www.medicalsuction.co.uk
7
MAINTENANCE
The NEW ASKIR 230/12V BR suction equipment does not need maintenance or lubrication.
It is, however to inspect the unit before each use. Unpack the instrument and always check integrity of plastic parts
and feeding cable, they might have been damaged during previous use. Connect cable to electrical network and turn
switch on.
Close the aspirator outlet with your finger and with suction regulator in maximum vacuum position check that the
vacuum indicators reaches
75 kPa (-0.75 bar) minimum (internal battery). Rotate the knob from right to left and check the aspiration regulating
control.
The vacuum indicator should go down 40 kPa (-0.40 bar). Verify that loud noises are not present, these can indicate
wrong functioning.
The equipment may also be used with the power cord hooked up to the electricity mains (the yellow light switches
over to the fixed green light).
Ensure that the green light comes on and that, after a few seconds it switches over to the yellow light (meaning that
the charged battery is operating).
A protection fuse (F 1 x 630mA 250V) reachable from exterior and it situated in the plug protects the instrument.
For fuse replacing, always the type and the range. The device is made up of a lead battery which cannot be accessed
by outside. In order to replace
consult the technical staff authorised by the manufacturer.
Fault type Cause Solution
1. Red light on Battery run down Hook up the power cord to the electricity mains,
positioning the
equipment power switch on 0. Charge it for 8
hours, until the
yellow light switch over to the green light.
2. Red light on Battery run down The device may be also be supplied by an power
cord hooked
up to the electricity mains: the yellow light comes
on and then
switch over to the green light.
3. No aspiration Jar Cap badly screwed down Unscrewed the cap, then rescrew it correctly
4. No aspiration Lid seal not in its seat Unscrew the cap and insert the seal properly in its
seat
5. The float doesn’t close If the cap has been washed,
ensure that the float is not
partially detached
Insert the float into it’s place
6. The float doesn’t close The float it’s covered by dirty
material
Unscrewed the cap, leave the and put in on
autoclave
7. Low suction Foam inside the jar Fill the jar to 1/3 full of ordinary water
8. No aspiration due to flow
leakage
of mucus
Filter blocked Replace filter
9. The Vacuum power on
the
patient side is either very
low or
absent
· Vacuum regulator set to
minimum
· Protection filter blocked or
damaged
· Connection tubes blocked,
kinked or disconnected
· Shut-off valve blocked or
damaged
· Pump motor damaged
· Turn the vacuum regulator clockwise and check
the value of
the vacuum on the gauge
· Replace the filter
· Replace or reconnect the tubes, check the jar
connections
· Empty the jar, or disconnect the tube from the
jar and
unblock the shut-off valve. The unit twill only work
in the
upright position
· Refer to authorised service personnel
Faults 1 - 2 - 3 - 4 - 5 - 6 -
7
None of the remedies has
achieved the desired results
Contact the seller or CA-MI After-sales Assistance
Service
If the overfill security system fails, continued aspiration will be stopped by the bacteriological filter.
Should this back-up security also fail, there is a risk of penetration of liquid into the device. Don’t attempt to proceed
with aspiration under these circumstances. Return the device to CA-MI for service.
7


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Cami New Askir 230 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Cami New Askir 230 in the language / languages: English as an attachment in your email.

The manual is 0,26 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info