Displayed temperature range 93.2 °F -108 °F (34 °C - 42.2 °C)
Operating ambient
temperature range 50 °F -104 °F (10 °C - 40 °C)
Display resolution 0.1 °F or °C
Accuracy for patient temperature range
Maximum Laborator
y Error
96.8 °F -102.2 °F (36 °C - 39 °C): ± 0.4 °F ± 0.2 °C
outside this range: ± 0.5 °F ± 0.3 °C
Long term storage ranges
Temperature: - 4 °F to 122 °F ( - 20 °C to 50 °C)
Humidity: 95 % non-condensing
Battery life: 2 years / 1000 measurements
This infrared thermometer meets requirements established in ASTM
Standard E 1965-98 (for the thermometer system [thermometer with
lens filter]). Full responsibility for the conformance of the product to the
standard is assumed by Braun GmbH, 61476 Kronberg, Germany.
ASTM laboratory accuracy requirements for the thermometer only (not
including a lens filter) in the display range of 98.8 °F to 102.2 °F
(36 °C to 39 °C) for infrared thermometers is ± 0.4 °F (± 0.2 °C), whereas
for mercury-in-glass and electronic thermometers, the requirement per
ASTM Standards E 667-86 and E 1112-86 is ± 0.2 °F ( ± 0.1 °C).
This appliance conforms to the following standards:
DIN EN 60601-1: 3/96 «Medical electrical equipment» –
Part 1: General requirements for safety
DIN EN 12470-5: 2003 «Clinical thermometers» –
Part 5: Performance of infrared ear thermometers
(with maximum device)
Internally Powered Equipment
Continuous Operation
Not Protected against Ingress of Water
U.S. Patent No. 5,088,834 Other Patents Pending
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC. For
detailed description of EMC requirements please contact the Braun ThermoScan Customer
Service at 1-800-327-7226.
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.
In the unlikely event you experience any difficulty using the Braun ThermoScan thermometer,
simply call us toll free at 1-800-327-7226.
15
TYPE BF
EQUIPMENT
Attention, consult
ACCOMPANYING
DOCUMENTS
MEDICAL ELECTRICAL EQUIPMENT
CLASSIFIED BY UNDERWRITERS LABORATORIES INC.
®
WITH RESPECT TO ELECTRIC SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH UL 2601-1 / CAN CSA C22.2
No.601.1 10HK
6012136_IRT_3000_NA 28.10.2005 8:20 Uhr Seite 15
15


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Braun Thermoscan Type 6012 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Braun Thermoscan Type 6012 in the language / languages: English as an attachment in your email.

The manual is 0,3 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Braun Thermoscan Type 6012

Braun Thermoscan Type 6012 User Manual - German, Dutch, French, Italian - 54 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info