Guidance and manufacturer’s declaration – electromagnetic emissions
The IRT 6020/6520 equipment is intended for use in the electromagnetic environment speci ed below.
The customer or the user of the IRT 6020/6520 should ensure that it is used in such an environment.
Emissions Test Compliance Electromagnetic environment – guidance
RF Emissions
CISPR 11
Group 1
The ME equipment uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF Emissions
CISPR 11
Class B Complies
Harmonic emissions
IEC 61000-3-2
Not
Applicable
The ME equipment is
solely battery powered.
Voltage  uctuations /
icker emissions
Not
Applicable
English
Non-Life Support Equipment Separation Distance Calculation
(3Vrms / 3V/m compliance)
Rated maximum output
power of transmitter (W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
in ISM bands
80 MHz to
800 MHz
800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.69 3.69 7.38
100 11.67 11.67 23.33
Guidance and manufacturer’s declaration – electromagnetic immunity
The IRT 6020/6520 is intended for use in the electromagnetic environment speci ed below.
The customer or the user of the IRT 6020/6520 should ensure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment – guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6kV Contact
±8kV Air
Complies
Floors should be wood,
concrete, or ceramic tile. If
oors are covered with
synthetic material, the relative
humidity should be at least
30%.
Radiated RF
IEC 61000-4-3
3 V/m 80MHz to
2.5GHz
Complies
Field strengths outside the
shielded location from  xed RF
transmitters, as determined by
an electromagnetic site survey,
should be less than 3 V/m.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Separation distance
calculation provided above. If a
known transmitter is present
the speci c distance can be
calculated using the equations.
Conducted RF
IEC 61000-4-6
3Vrms 150kHz to
80MHz
Not Applicable
(no electrical
cabling)
Electrical fast transient
IEC 61000-4-4
±2kV power line
±1kV I/O lines
Not Applicable
The ME equipment
is solely battery powered.
Surge
IEC 61000-4-5
±1kV
di erential
±2kV common
Not Applicable
Power frequency magnetic
eld
IEC 61000-4-8
3 A/m Complies
Power frequency
magnetic  elds should be at
levels characteristic of a
typical location in a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
>95% dip 0.5 cycle
60% dip 5 cycles
70% dip 25 cycles
95% dip 5 sec.
Not Applicable
The ME equipment
is solely battery
powered.
d = [
3,5
] P
V
1
d = [
3,5
] P
E
1
d = [
7
] P
E
1
8


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Braun ThermoScan 5 - IRT6020 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Braun ThermoScan 5 - IRT6020 in the language / languages: English, German, Dutch, French, Italian, Portuguese, Portuguese, Spanish as an attachment in your email.

The manual is 9,59 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info