41741
11
Zoom out
Zoom in
Previous page
1/63
Next page
11
Troubleshooting
Error message
Situation Solution
No probe cover is attached
and activation button was
pressed while probe was in the
ear.
Make sure probe window is clean;
refer to section «Care and cleaning»
if necessary.
Attach new, clean probe cover to
stop error beeps.
Ambient temperature is not
within the allowed operating
range of 10°C-40°C
(50°F to 104°F) or changing
too rapidly.
Allow the thermometer to remain in a
room for 30 minutes where
the temperature is between
10°C and 40°C (50°F to 104°F).
Temperature taken is not
within displayed temperature
range (20°C-42.2°C/ 68°F-
108°F).
HI = too high
LO = too low
Make sure new, clean probe cover is
attached and thermometer is
properly inserted.
To clear the error message, attach a
new, clean probe cover or press the
/mem
button.
Then, take a new temperature.
Battery is low, but
thermometer will still operate
correctly.
(No light function)
Insert new batteries
Battery is too low to take
correct temperature readings.
Insert new batteries
System error
• If error persists
• If error still persists
Wait 30 seconds until thermometer
powers down, then turn on again.
… reset the thermometer by
removing the batteries and putting
them back in.
…call Welch Allyn Customer Service
(see page 2).
6014126_WA_MN Seite 11 Montag, 12. November 2001 7:22 07
11


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Braun Pro3000 Thermoscan at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Braun Pro3000 Thermoscan in the language / languages: English, German, Dutch, French, Italian, Spanish as an attachment in your email.

The manual is 1,4 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info