9
English
Errors and troubleshooting
Error message Situation Solution
No lens lter is attached. Attach new, clean lens lter.
The thermometer probe was not
positioned securely in the ear.
An accurate measurement
was not possible.
POS = position error
Check if the positioning of the probe is
correct and remains stable.
Replace lens lter and reposition.
Press Start button to begin a new
measurement.
Ambient temperature is not within
the allowed operating range
(50 – 104 °F or 10 – 40 °C).
Allow the thermometer to remain for
30 minutes in a room where
the temperature is between
50 and 104 °F or 10 and 40 °C.
Temperature taken is not within
typical human temperature range
(93.2 – 108 °F or 34 – 42.2 °C).
HI = too high
LO = too low
Make sure the probe tip and lens are
clean and a new, clean lens lter is
attached. Make sure the thermometer
is properly inserted. Then, take a new
temperature.
System error – self-check display
ashes continuously and will not be
followed by the ready beep and the
ready symbol or screen is blank.
If error persists,
If error persists,
If error still persists,
Wait 1 minute until the thermometer
turns o automatically, then turn
on again.
... reset the thermometer by removing
the batteries and putting them back in.
... replace the batteries with new
batteries.
... please contact the Service Center.
Battery is low, but thermometer will
still operate correctly.
Insert new batteries.
Battery is too low to take correct
temperature measurement.
Insert new batteries.
11


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Braun IRT 6515 ThermoScan at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Braun IRT 6515 ThermoScan in the language / languages: English, Spanish as an attachment in your email.

The manual is 4,08 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Braun IRT 6515 ThermoScan

Braun IRT 6515 ThermoScan User Manual - English, German, Dutch, French, Italian, Portuguese, Turkish, Spanish - 37 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info