13
This product conforms to the provisions of the EC directive 93/42/EEC (Medical Device
Directive). This device conforms to the following standards:
EN 60601-1: 2006 + AC:2010: - General requirements for basic safety and essential
performance
EN 60601-1-2:2007 - Electromagnetic compatibility requirements and tests
EN 60601-1-11:2010 - Requirements for medical electrical equipment and medical
electrical systems used in the home healthcare environment
EN 1060-1:1995 + A2:2009 – Non-invasive sphygmomanometers - general
requirements
EN 1060-3:1997 + A2:2009 - Non-invasive sphygmomanometers - Supplementary
requirements for electro-mechanical blood pressure measuring systems.
EN 1060-4:2004 - Non-invasive sphygmomanometers - Test procedures to determine
the overall system accuracy of automated non-invasive sphygmomanometers.
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC.
For detailed description of EMC requirements please contact an authorized local Service Centre (see
package insert).
Portable and mobile RF communications equipment can aect medical electrical equipment.
Please do not dispose of the product in the household waste at the end of its useful life.
Disposal can take place at your local retailer or at appropriate collection points provided
in your country.
Guarantee
See warranty statement inside the Guarantee/Consumer Card/Authorized Service Centers booklet.
If for any reason you have to return your device to the point of sales or authorized Service Center,
please fully complete the Consumer Card (provided with the device) and insert it inside of the box
before you return your device.
The Consumer Card can also be found in Kaz Europe website (www.hot-europe.com/customer-care/
repair-services/) and you can download it in the case that you lose the one provided with the device.
The LOT and SN of your device are printed on the rating label in the back of the product.
UK only:
This guarantee in no way aects your rights under statutory law.
0297
Specications
Method of measurement Oscillometric
Model number BP6000, BP6100, BP6200
Range of measurement Pressure 0~300 mmHg
Pulse 40~199 beats/minute
Accuracy Pressure +/- 3 mmHg
Pulse +/- 5 % Max.
Ination Deluxe automatic
Display Liquid crystal display – systolic, diastolic, pulse rate
Backlight Display for BP6200
Sets of memory BP6000: 40 sets per user
BP6100: 50 sets per user
BP6200: 60 sets per user
Cu size Small cu = 22-32cm arm circumference
Large cu = 32-42cm arm circumference
Operating temperature +10 °C ~ + 40 °C, less than 85 % R.H.
Storage temperature -20 °C ~ +60 °C, less than 85 % R.H.
Unit weight Approximately 500gr (without batteries)
Power supply Alkaline battery: 4 x AA (LR6)
Battery life 300 times measurement
Auto power o Whenever not used for 1 minute
Accessories 4 batteries, 2 arm cus with tube, instruction manual, pouch,
travel bag
IMPORTANT
Read the operating instructions.
Classication:
Internally powered equipment
Type B equipment
IP22
Not suitable for use in the presence of ammable anesthetic mixture with air, oxygen
or nitrous oxide
Continuous operation with short-time loading
Operating temperature
Storage temperature
7


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Braun ExactFit 5 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Braun ExactFit 5 in the language / languages: English, German, Dutch, Danish, French, Italian, Polish, Portuguese, Swedish, Turkish, Spanish, Norwegian, Finnish as an attachment in your email.

The manual is 34,7 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info