445336
63
Zoom out
Zoom in
Previous page
1/92
Next page
63
Caratteristiche tecniche
Modello: BP 1750 / BP1700
Principio operativo: Metodo oscillometrico
Display: Cristalli liquidi
Intervallo di misurazione: 0-320 mmHg (pressione)
40-199/min (frequenza cardiaca)
Misurazione della
pressione arteriosa: 20 mmHg (minimo valore diastolico)
300 mmHg (massimo valore diastolico)
Accuratezza di laboratorio: ±3 mmHg (pressione)
±5% valore rilevato (frequenza cardiaca)
Accuratezza clinica: Secondo AAMI-SP10 con metodo auscultatorio
di riferimento:
< 5 mmHg variazione sistematica
< 8 mmHg deviazione standard
Gonfiaggio: Automatico
Sgonfiaggio: Valvola graduale
Alimentazione: 2 batterie di tipo LR 03 (AAA)
Temperatura operativa: da + 10 a + 40 °C
Temperatura di conservazione: da – 20 a +60 °C
Umidità relativa: Fino al 65%
Manicotto: Adeguato per circonferenza del polso
tra 13 e 21 cm
Taratura clinica: Secondo AAMI-SP 10 con metodo auscultatorio
di riferimento.
(Notate che una valutazione nei confronti di
misurazioni tra arterie può determinare risultati
diversi per apparecchi che sono tarati rispetto al
metodo auscultatorio di riferimento.)
Se lo strumento non è utilizzato all'interno delle tolleranze di temperatura ed umidità
specificate non è possibile garantirne la precisione tecnica.
Dispositivo tipo BF
Soggetto a variazioni senza preavviso.
6075360_BP1700_KAZ Seite 63 Donnerstag, 19. Juli 2007 8:35 08
63


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Braun BP 1750 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Braun BP 1750 in the language / languages: English, German, Dutch, French, Italian, Portuguese, Spanish as an attachment in your email.

The manual is 2,44 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info