Product specifications
Type: No touch & touch forehead thermometer (BNT 400)
Measuring range: 93.2 °F – 109.4 °F (34 °C – 43 °C)
Resolution: 0.1 °F (0.1 °C)
Laboratory Accuracy: 0.4 °F for 95 °F – 107.6 °F (0.2 °C for the range 35 °C – 42 °C)
outside that range ±0.5°F/0.3°C
(Ambient Temperature: 59 °F to 104 °F (15 °C to 40 °C)
This thermometer displays a calculated oral equivalent estimate.
Display: Liquid Crystal Display, 4 digits plus special icons
Acoustic: Audio:
Normal temp range = Green temp range: 1 long beep for 0.4 second duration
Fever = Red or Yellow temp range: 10 short beeps for 0.2 second duration
Operating temperature:
59
°
F - 104
°
F ( 15
°
C - 40
°
C )
Automatic Switch-o: Approx. 30 seconds after last measurement has been taken
Weight: 3.54 oz. (with batteries), 2.91 oz. (w/o batteries)
Long term storage ranges
Storage/transport temperature: -13 °F – 131 °F (-25 °C to 55 °C)
Humidity: 15–95% non-condensing
Battery: (2) AAA Batteries - at least 500 measurements
Pressure: 700-1060 hPA (0.7-1.06 atm)
This infrared thermometer meets requirements established in ASTM Standard E 1965-98 (for the
thermometer system). Full responsibility for the conformance of the product to the standard is
assumed by Kaz USA, Inc., a Helen of Troy Company, Health & Home, 400 Donald Lynch Blvd., Suite 300,
Marlborough, MA 01752
ASTM laboratory accuracy requirements for the thermometer only in the display range of 36 °C to 39°C
(96.8 °F to 102.2 °F) for infrared thermometers is ±0.3 °C (±0.5 °F), whereas for mercury-in-glass and
electronic thermometers, the requirement per ASTM Standards E 667-86 and E 1112-86 is ±0.1 °C (±0.2 °F).
This device conforms to the following standards:
EN 60601-1:2014 Medical electrical equipment. General requirements for basic safety and
essential performance
ASTM E1965-98:2016 - Standard Specication for Infrared Thermometers for Intermittent Determination
of Patient Temperature
ISO 80601-2-56:2017 Particular requirements for basic safety and essential performance of clinical
thermometers for body temperature measurement
IEC 60601-1-2:2014: Medical electrical equipment – Part 1-2: General requirements for basic safety and
essential performance – Collateral Standard: Electromagnetic disturbances – Requirement and tests
NOTE: Do not use this device in the presence of electromagnetic or other interference outside the normal
range specied in IEC 60601-1-2:2007.
EN 1041: 2008 Information supplied by the manufacturer of medical devices
IEC 60601-1-11: 2015 Medical electrical equipment – Part 1-11: General requirements for basic safety and
essential performance – Collateral standard: Requirements for medical electrical equipment and medical
electrical systems used in the home healthcare environment
English
15


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Braun BNT400 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Braun BNT400 in the language / languages: English, Spanish as an attachment in your email.

The manual is 4,1 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info