Specifications
Model: BP 1750 / BP 1700
Operation principle: Oscillometric method
Display: Liquid Crystal Display
Measuring range: 0–320 mmHg (cuff pressure)
40–199 / min (pulse rate)
Blood pressure
measurement: 20 mmHg (minimum diastolic value)
300 mmHg (maximum systolic value)
Laboratory accuracy: ± 3 mmHg (cuff pressure)
± 5 % of reading (pulse rate)
Clinical accuracy: According to AAMI-SP10 with auscultatory reference:
< 5 mmHg systematic offset
< 8 mmHg standard deviation
Inflation: Automatic
Deflation: Step valve
Power supply: 2 batteries, type LR 03 (AAA)
Operating temperature: +10 °C to + 40 °C
Storage temperature: – 20 °C to + 60 °C
Humidity: Up to 85% RH
Cuff: Suitable for wrist circumference of 13 to 21 cm
Clinical validation: According to AAMI-SP10 with auscultatory reference.
(Please note that a validation against intra-arterial
measurements can lead to different results for devices
which are calibrated against auscultatory reference.)
If device is not used within specified temperature and humidity ranges the technical
accuracy can not be ensured.
Type BF equipment
Subject to change without notice.
This device conforms to the following standards:
DIN EN 60601-1: 3/96 «Medical electrical equipment» –
Part 1: General requirements for safety
6075360_BP1700_KAZ Seite 22 Donnerstag, 19. Juli 2007 8:35 08
22


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Braun 6076 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Braun 6076 in the language / languages: English, German, Dutch, French, Italian, Portuguese, Spanish as an attachment in your email.

The manual is 2,44 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info