11
Problem Reason Solution
e.g. Error message lower than
64:
System error
Have the product checked at a
Braun Service Centre.
Measurement or
inflation does not
start instantly.
Right measurement
position was not reached.
Repeat measurement as shown
and described under «Measuring»
or deactivate the position sensor.
Readings from
repeated mea-
surements differ
considerably.
Blood pressure is a
fluctuating value.
For healthy adults,
deviations of 10 to
20 mmHg are possible.
none
Cuff is not properly fitted. Ensure that the appliance is fitted
as follows:
on top of the slim side of the
forearm
with the display on top
with the battery compartment
facing the users body
approx. 2 cm away from the
thumb root
Readings were not con-
sistently taken at heart
level.
For each measurement, proceed
exactly as shown and described
under «Measuring».
Talking, coughing,
laughing, moving etc.
when measuring will
influence the reading.
While measuring, relax, keep still,
do not move, or speak.
Prior to a mea-
surement, the
position sensor
icon on the display
does not work.
Position sensor is de-
activated.
Refer to the chapter «Settings»
to activate the position sensor.
6063010_BP2550_S4-58 Seite 11 Dienstag, 8. April 2003 2:17 14
11


Need help? Post your question in this forum.

Forumrules
1

Forum

braun-6053-bp2550-sensorcontrol

Reset search

  • Tensiómetro 6053.: necesito service, probable balón pinchado.
    Reg. N*1114 E98 SSL , CE0297-LOTE 1.169 Ow 1920 xo ol 6H - Submitted on 30-11-2021 at 13:46

    Reply Report abuse


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Braun 6053 BP2550 SensorControl at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Braun 6053 BP2550 SensorControl in the language / languages: English, Dutch, French, Portuguese, Spanish as an attachment in your email.

The manual is 1,15 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info