533560
137
Zoom out
Zoom in
Previous page
1/140
Next page
GB
Non-contact fever thermometer JFT 90
137
Guidance and manufacturer’s declaration - electromagnetic immunity
The JFT90 is intended for use in the electromagnetic environment specified below. The customer or the
user of the JFT90 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic
environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Electrical fast transient/
burst
IEC 61000-4-4
2 kV for power supply
lines
1 kV for input/output lines
Not applicable
Mains power quality
should be that of a typi-
cal commercial or
hospital environment.
Surge
IEC 61000-4-5
1 kV line(s) to line(s)
2 kV line(s) to earth
Not applicable
Mains power quality
should be that of a typi-
cal commercial or
hospital environment.
137


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Beurer JFT90 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Beurer JFT90 in the language / languages: English, German, French, Italian, Polish, Turkish, Spanish as an attachment in your email.

The manual is 2,45 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Beurer JFT90

Beurer JFT90 User Manual - Dutch - 20 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info