bacterial growth and to ensure effective therapeutic
use, this moisture must always be removed. To remove
moisture from the hose, proceed as follows:
Remove the hose [7] from the atomiser [8].
Leave the hose plugged into the nebuliser [4].
Run the nebuliser until all the moisture has been dis-
placed by the flow of air.
In the event of significant contamination, replace the
hose.
Disinfection
Follow the steps listed below carefully in order to disin-
fect your atomiser and the accessories. It is advisable
to disinfect the individual components after the last
time they are used each day at the latest.
First clean the atomiser and the accessories as de-
scribed under “Cleaning”.
The atomiser should be taken apart and placed to-
gether with the mouthpiece and nosepiece in boiling
hot water for at least 10 minutes. Avoid the parts
coming into contact with the hot base of the pot.
You can also disinfect the atomiser, the mouthpiece
and the nosepiece with a standard vaporizer. Ob-
serve the manufacturer’s instructions for use for the
vaporizer when doing this.
Note
The face masks and air hose must not be placed in
hot water!
Carefully dry all components using a soft cloth.
We recommend disinfecting the masks using a retail
disinfectant.
Once all components are fully dry, reassemble the
atomiser and place the components in a dry, sealed
container.
Note
Ensure that all parts are thoroughly dried during, and
in particular after cleaning, otherwise there is an in-
creased risk of bacterial growth.
Drying
Place the individual components on a dry, clean and
absorbent surface and allow to dry completely (for
at least 4 hours).
Material resistance
As with all plastic components, frequent use and
cleaning of the atomiser and the accessories can lead
to a certain amount of wear. Over time, this can alter
the aerosol properties and may eventually affect the
efficiency of therapy. We therefore recommend that
you replace the atomiser after no more than a year.
When choosing a cleaning or disinfecting agent, note
the following: Only use mild cleaning or disinfect-
ant agents in the quantities recommended by the
manufacturer.
Storage
Do not store the nebuliser in a damp atmosphere
(e.g. in the bathroom) or transport it together with
damp objects.
Store and transport the nebuliser away from direct
sunlight.
Storage
The device and all accessories, e.g. the nebuliser
and masks, can be stored together in the storage
bag provided.
9. Disposal
In the interest of environmental protection, do not dis-
pose of the device in household waste.
Please dispose of the device in accordance with
EC Directive – WEEE (Waste Electrical and Elec-
tronic Equipment).
If you have any queries, please contact the ap-
propriate local authorities.
10. Troubleshooting
Problem/
Question
Possible Cause/Remedy
The atomiser
produces
little or no
aerosol.
1. Too much or too little medicine in the
atomiser.
Minimum: 0.5 ml, Maximum: 8 ml.
2. Check that the nozzles are not
blocked If necessary, clean the noz-
zle (e.g. by rinsing). Use the atomiser
again.
CAUTION: Only poke implements
through the fine holes from the
underside of the nozzle.
3. The atomiser is not held in an upright
position.
4. The medicine solution is unsuitable
for nebulising (e.g. too thick).
The medicine solution should be
specified by the doctor.
Output is too
low.
Hose kinked, filter blocked, too much
inhalation solution.
Which
medicines are
suitable for
inhalation?
Consult your doctor.
Essentially all medicines that are suit-
able and approved for device-based
inhalation can be inhaled.
Some inhala-
tion solution
remains in the
atomiser.
This is normal and occurs for technical
reasons. Stop inhalation if you hear a
notable difference in the noise of the
atomiser.
15


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Beurer IH 25-1 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Beurer IH 25-1 in the language / languages: English, German, French, Italian, Polish, Turkish, Spanish as an attachment in your email.

The manual is 5,3 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Beurer IH 25-1

Beurer IH 25-1 User Manual - Dutch - 12 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info