36 Beurer GL50 mmol/L
Only a small amount of blood is required (0.6 microlitres) and measurement takes approx. five sec-
onds. The test strip detects blood glucose values from 1.1 to 35.0 mmol/L.
Chemical components of the test strip sensor
FAD glucose dehydrogenase 6%
Potassium ferricyanide 56%
Non-reactive components 38%
Control solution functionality
The control solution contains a fixed amount of glucose that reacts with the test strip. A test with
control solution is similar to a blood test. However, control solution is used instead of blood. The
measured result using control solution must be within the result range. This range is printed on each
test strip box.
Chemical composition of the control solution
The control solution is a red solution with the following D-glucose level (in percentage shares):
Ingredients Control solution LEVEL 3 Control solution LEVEL 4
D-glucose 0.14% 0.37%
Non-reactive components 99.86% 99.63%
Standards
The Beurer GL50 blood glucose monitor complies with the European standards: IVD (98/79/EC) and
MDD (93/42/EC).
Comparison of measured values with laboratory values
Performance characteristics: accuracy and precision
Whole blood glucose test results were compared with the YSI 2300 laboratory device. At a concentra-
tion of < 4.2 mmol/L ≥ 98% were at +/-0.8 mmol/L, while at a glucose concentration of ≥ 4.2 mmol/L
98% were within 20% of the reference values. The CV (coecient of variation) is < 5%. The blood
glucose monitor is therefore comparable to a laboratory system.
You will find further details and information regarding blood glucose results and various technologies
in general medical literature on the subject.
36


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Beurer GL 50 mmol L at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Beurer GL 50 mmol L in the language / languages: English as an attachment in your email.

The manual is 3,46 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Beurer GL 50 mmol L

Beurer GL 50 mmol L User Manual - German - 40 pages

Beurer GL 50 mmol L User Manual - Dutch - 40 pages

Beurer GL 50 mmol L User Manual - French - 40 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info