26 Beurer GL50 mmol/L
4
Drip control solution on surface
Choose a clean surface to carry out the
functional test correctly. Shake the con-
trol solution well before use. Undo the
cap and press two drops next to each
other on the clean surface without touch-
ing them. Use the second drop for the
measurement.
Note
Never apply the control solution straight from the bottle to a test strip. Reason: The remain-
ing solution in the bottle will be contaminated if the top of the bottle comes into contact with
the test strip.
5
Apply drop to the test strip
Hold the input gap (at the tip of the test strip) to the drop of control solution until the gap
is completely filled. When the gap is suciently filled with solution, the device performs a
measurement. The device counts down for approx. five seconds. The measured result is then
shown on the display.
6
Evaluate result of functional test
Check whether the result is within the specified range of results for the control solution. This
range of results is printed on the test strip box.
Expected results
At room temperature, the measured values from the test using the control solution should be within
the range printed on the test strip box in approx. 95% of all tests.
Warning
The result range printed on the test strip box applies only to the control solution. This is not a rec-
ommended value for your blood glucose level.
26


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Beurer GL 50 mmol L at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Beurer GL 50 mmol L in the language / languages: English as an attachment in your email.

The manual is 3,46 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Beurer GL 50 mmol L

Beurer GL 50 mmol L User Manual - German - 40 pages

Beurer GL 50 mmol L User Manual - Dutch - 40 pages

Beurer GL 50 mmol L User Manual - French - 40 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info