537054
35
Zoom out
Zoom in
Previous page
1/40
Next page
Beurer GL50 mg/dL 35
9 technIsche gegevens
Afmetingen (L x B x H) 123 x 28 x 16 mm
Gewicht 36 g (inclusief batterijen)
Voeding 2 x 3 V CR 2032 knoopcellen
Levensduur batterijen
Meer dan 1000 metingen
Meetwaardengeheugen 480 meetwaarden met datum/tijd
Gegevensbehoud bij vervangen van batterijen
Gemiddelde waarden over 7, 14, 30, 90 dagen
Automatische uitschakeling 2 minuten na laatste bediening
Temperatuur opslag/
vervoer
Temperatuur: +2 °C – +30 °C
Relatieve luchtvochtigheid: < 90%
Werkingsbereiken Temperatuur: +10 °C – +40 °C
Relatieve luchtvochtigheid: < 90% niet condenserend
Meetbereik glucose Glucose: 20 – 630 mg/dl
Bloedmonster capillair volbloed
Vereiste hoeveelheid bloed 0,6 microliter
Meetduur bloedsuiker ongeveer 5 seconden
Kalibreren Plasma
Testprocedure Amperometrische biosensor
Gebruik Geschikt voor thuisgebruik
Systeemfunctietest Bij elke keer inschakelen
EMC
Dit apparaat voldoet aan de Europese norm EN 61326 en is onderworpen aan bijzondere veiligheids-
maatregelen op het gebied van elektromagnetische compatibiliteit. Let er daarbij op dat draagbare en
mobiele HF-communicatie-installaties dit apparaat kunnen beïnvloeden. U kunt uitgebreide informatie
aanvragen bij de klantenservice op het aangegeven adres.
Over de werking van de teststrips
De teststrips maken een meting van de hoeveelheid glucose in vers volbloed mogelijk. Wanneer de
gleuf voor bloedopname in contact komt met een druppel bloed, vult hij zich automatisch door een-
voudige capillaire werking. Het bloed wordt in de absorberende gleuf van de teststrip gezogen en
het meetapparaat meet de suikerspiegel van het bloed.
De test bestaat uit het meten van een elektrische spanning die wordt veroorzaakt door de chemische
reactie tussen het enzym glucose-dehydrogenase (Aspergillus sp.) en het reagens van de strip.
Tijdens de reactie transporteert een mediator elektronen door het elektrodenoppervlak en genereert
hierdoor een stroom.
Het meetapparaat meet deze spanning. De spanning is evenredig aan het glucosegehalte van het
bloedmonster. Het resultaat wordt op het display van het bloedsuikermeetapparaat weergegeven.
35


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Beurer GL 50 mg dL at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Beurer GL 50 mg dL in the language / languages: Dutch as an attachment in your email.

The manual is 3,34 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Beurer GL 50 mg dL

Beurer GL 50 mg dL User Manual - English - 40 pages

Beurer GL 50 mg dL User Manual - German - 40 pages

Beurer GL 50 mg dL User Manual - French - 40 pages

Beurer GL 50 mg dL User Manual - Italian - 40 pages

Beurer GL 50 mg dL User Manual - Spanish - 40 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info