The battery symbol is shown. How to proceed:
Replace all batteries.
You have an unpleasant sensation at the electrodes. How to
proceed:
(1)
The electrodes are not positioned correctly. Check their positions
and re-position, if necessary.
(2)
The electrodes are worn. This may cause irritated skin as even
distribution of the current across the entire area is no longer
guaranteed. For this reason, replace the electrodes.
Skin in the treatment area turns red. How to proceed:
Immediately stop treatment and wait until your skin has returned to its
normal condition. If the redness is under the electrode and disappears
quickly, there is no risk – this is caused by the locally stimulated,
increased blood flow.
However, consult your doctor before you continue treatment if the
skin irritation persists and this is accompanied by an itchy sensation
or inflammation. This may be caused by an allergic reaction to the
adhesive surface.
13. Replacement parts and wearing parts
You can obtain the following replacement parts directly from
Customer Services:
Designation Item number and/or order number
8 x adhesive electrodes
(45 x 45 mm)
Item 661.02
4 x adhesive electrodes
(50 x 100 mm)
Item 661.01
14. Technical specifications
Name and model EM 49
Type EM 49
Output waveform Biphasic rectangular pulse
Pulse length 50 – 450 µs
Pulse frequency 1 – 150 Hz
Output voltage max. 100 Vpp (500 ohm)
Output current max. 200 mApp (500 ohm)
Voltage supply 3 x AAA batteries
Treatment time Adjustable from 5 to 100 minutes
Intensity Adjustable from 0 to 50
Operating conditions 5°C – 40°C (41°F – 104°F) at a relative
humidity of 15 – 93%
Storage conditions -25 °C –70 °C (-13 °F – 158 °F) at a
relative humidity of 0– 93%
Dimensions 132 x 63 x 29.5 mm (including belt clip)
Weight
83 g (including belt clip, without batteries),
117 g (including belt clip and batteries)
Use Altitude Limit 3000 m
Use Range Atmospheric
Pressure
700 - 1060 hPa
The serial number is located on the device or in the battery
compartment.
Note: If the device is not used acording to the instructions specified,
perfect functionality cannot be guaranteed!
We reserve the right to make technical changes to improve and
develop the product.
This device complies with European standards EN60601-1 and
EN60601-1-2 (In accordance with IEC 61000-4-2, IEC 61000-4-3,
IEC 61000-4-4, IEC 61000-4-5, IEC 61000-4-6, IEC 610004-8 and
IEC 610004-11) and is subject to special precautionary measures with
regard to electromagnetic compatibility. Please note that portable and
mobile HF communication systems may interfere with this unit.
More details can be requested from the stated Customer Services
address or found at the end of the instructions for use.
36


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Beurer EM 49 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Beurer EM 49 in the language / languages: English, German, French, Italian, Polish, Turkish, Spanish as an attachment in your email.

The manual is 17,88 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Beurer EM 49

Beurer EM 49 User Manual - Swedish - 20 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info