577368
54
Zoom out
Zoom in
Previous page
1/58
Next page
08/ 2012 54 / 58
BENNING ST 750 A
20.2 Factory Settings, Limiting Values Setup
Themeaningofthedataisdescribedinchapter7.1"LimitingValues"!
RPE data
RPE VDE 0701-0702
RPE VDE 0751-1
RPE per 7.5 m line length
Line length
Cross-section
(Ω)0.3
(Ω)0.3
(Ω)0.1
(m) 5
(mm²) 1.5
IAL data
IPE VDE 0701-0702 Class I
ILeak VDE 0751 Class I
ILeak VDE 0751 Class II
ILeak VDE 0751 note 1
ILeak VDE 0751 note 2
ILeakVDE0751X-raywithPE
ILeakVDE0751X-raywithoutPE
Patient leakage current BF
Patient leakage current CF
(mA) 3.5
(mA) 1.0
(mA) 0.5
(mA) 5.0
(mA) 10.0
(mA) 5.0
(mA) 2.0
(mA) 5.0
(mA) 0.05
ICont data
ICont VDE 0701-0702 (mA) 0.5
Ua data
Contact voltage U to PE (V) 25
RInsu data
RInsu Class I
RInsu Class I heating element
RInsu Class II
RInsu Class III
RInsu 0751 Class I
RInsu 0751 Class II
RInsu 0751 CF
Testing voltage prim. - PE
Testing voltage sec. - PE
Testing voltage prim. - sec.
(MΩ)1.0
(MΩ)0.3
(MΩ)2.0
(MΩ)0.25
(MΩ)2.0
(MΩ)7.0
(MΩ)2.0
(V) 500
(V) 500
(V) 500
ILeak data
IPE VDE 0701-0702 Class I
ILeak VDE 0751 Class I
ILeak VDE 0751 Class II
ILeak VDE 0751 note 1/3
ILeak VDE 0751 note 2
Patient leakage current BF (AC)
Patient leakage current CF (AC)
(mA) 3.5
(mA) 0.5
(mA) 0.1
(mA) 2.5
(mA) 5.0
(mA) 0.1
(mA) 0.01
Funct. data
IFault VDE 0701-0702 Class I
IFault VDE 0701-0702 Class II
IFault VDE 0751 Class I
IFault VDE 0751 Class II
(mA) 3.5
(mA) 0.5
(mA) 0.5
(mA) 0.1
Cable data
Line length
Cross-section
Conductor quantity
(m) 5
(mm
2
) 1.5
3
54


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning ST 750A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning ST 750A in the language / languages: English as an attachment in your email.

The manual is 7,87 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Benning ST 750A

Benning ST 750A User Manual - German - 68 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info