577368
50
Zoom out
Zoom in
Previous page
1/58
Next page
08/ 2012 50 / 58
BENNING ST 750 A
Measurement acc. to measuring spec. / limiting value Tester acc. to regulation
Measurement of the alternative leakage current,
as protective conductor current, contact current ,
device leakage current, leakage current of applied part for medical
electrical devices (patient leakage current)
DIN VDE 0701-0702, Class I
Protective conductor current, 3.5 mA
Heating devices up to 3.5 kW 3.5 mA
for > 3.5 kW 1 mA / kW,
max. 10.0 mA
DIN VDE 0701-0702
Contact current (ICont)
Classes I - II 0.5 mA
DIN VDE 0751-1
Class I ILeak, 1.0 mA
Class II ILeak, 0.5 mA
DIN VDE 0751-1
PLeak (AC) type BF 5.00 mA
type CF 0.05 mA
IEC 60601-1
ILeak note 1 5.00 mA
ILeak note 2 10.00 mA
ILeakX-raywithPE 5.00mA
ILeakX-raywithoutPE 2.00mA
DIN EN 61010-1
(VDE 0411-T1)
DIN EN 61557-1
(VDE 0413 T1)
DIN VDE 0404-1
DIN VDE 0404-2
Testing voltage: approx. 200 V AC, ± 20% at 230 V AC
The measured current is projected to 230V!
Testing current: max. 25 mA
Measuring range: 0.050 ... 25.000 mA
Resolution: 0.001 mA
Accuracy: at 230 V AC ± 10%
± 15% of the measured value
Internal resistance (measuring instrument / probe)
VDE0701–0702 Ri 2kΩ
VDE0751-1 Ri 1kΩ
Measurement acc. to measuring spec. / limiting value Tester acc. to regulation
Measurement by means of the differential current measurement method
(test socket); measurement of the protective conductor current for
devices of Class I (test socket), measurement of the leakage current,
contact current, leakage currents of applied parts for medical electrical
devices, patient leakage current; protection classes I and II
DIN VDE 0701-0702, Class I
Protective conductor current, 3.5 mA
Heating devices up to 3.5 kW 3.5 mA
for > 3.5 kW 1 mA / kW,
max. 10.0 mA
DIN VDE 0701-0702
Contact current (ICont)
Classes I - II 0.5 mA
DIN VDE 0751-1
Class I ILeak, 0.5 mA
Class II ILeak, 0.1 mA
DIN EN 61010-1
(VDE 0411-T1)
DIN EN 61557-1
(VDE 0413 T1)
DIN VDE 0404-1
DIN VDE 0404-2
DIN VDE 0404-3
Measuring range: 0.050 ... 25.000 mA (displayed from 50 µA on)
Resolution: 0.001 mA
Accuracy: at 230 V AC ± 10%
± 15% of the measured value (from 100 µA on)
manual / automatic pole reversal!
Internal resistance (measuring instrument / probe)
VDE0701–0702 Ri 2kΩ
VDE0751-1 Ri 1kΩ
50


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning ST 750A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning ST 750A in the language / languages: English as an attachment in your email.

The manual is 7,87 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Benning ST 750A

Benning ST 750A User Manual - German - 68 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info