9. Overview for Tests according to DIN VDE 0751-1 / EN 62353:2008-08
Protection class I
(with protective conductor)
Protection class II
(without protective conductor)
5.1* Scope of testing: *) The gures stated refer to DIN VDE 0751-1:2008-08.
5.2 Visual inspection / inspection
Check for visible damages of the following:
- connecting cables / plug-and-socket connections
- housing
- strain relief, protection against bending and kinking ...
5.3.2 Measuring the protective conductor resistance
Continuity between the earthing contact of the mains plug and accessible con-
ductive parts of the device or of the device connection
Medical electrical device with non-detachable mains connection cable: ≤ 0.3 Ω
Medical electrical device with detachable mains connection cable: ≤ 0.2 Ω
Detachable mains connection cable: ≤ 0.1 Ω
Detachable mains connection cable + medical electrical device: ≤ 0.3 Ω
Medical electrical device with permanent connection: ≤ 0.3 Ω
System with multiple socket: ≤ 0.5 Ω
5.3.3 Measuring leakage currents:
5.3.3.2 Device leakage current
Alternative leakage current measurement:
Direct / differential current measurement:
5.3.3.3 Leakage current of applied part
Alternative leakage current measurement (AC*):
Direct current measurement (AC*):
Type B
≤ 1000 µA
≤ 500 µA
---
---
Type BF
≤ 1000 µA
≤ 500 µA
≤ 5000 µA
≤ 5000 µA
Type CF
≤ 1000 µA
≤ 500 µA
≤ 50 µA
≤ 50 µA
Type B
≤ 500 µA
≤ 100 µA
---
---
Type BF
≤ 500 µA
≤ 100 µA
≤ 5000 µA
≤ 5000 µA
Type CF
≤ 500 µA
≤ 100 µA
≤ 50 µA
≤ 50 µA
5.3.4 Measuring the insulating resistance
No limiting values in VDE 0751-1 / EN 62353 standard.
No limiting values in VDE 0751-1 / EN 62353 standard.
5.4 Functional test
Functioning of safety equipment and function test
6. Measuring results, analysis, evaluation, documentation
(inspection and preparation for intended use)
* Note 1: The DIN EN 62353 (VDE 0751-1):2008-08 standard does not contain any measuring methods and admissible values for devices generating DC leakage currents. In this case, the
manufacturer should provide specifications in the accompanying documents.
* Note2:"Particularrequirements"mightallowotherleakagecurrentvalues.
* Note 3: In the current issue of the DIN VDE 0751-1 standard, the measurement of the insulating resistance is required, if it is considered to be convenient. Insulation measurement must not be carried
out, if it is excluded in the accompanying documents according to the manufacturer's specifications. Limiting values for the insulating resistance are not specified in the DIN VDE 0751-1/ EN 62353
standard (issue 2008). Here, it is possible to refer to recommendations of the manufacturer or to previously measured values.
17


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning ST 750A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning ST 750A in the language / languages: English as an attachment in your email.

The manual is 7,87 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Benning ST 750A

Benning ST 750A User Manual - German - 68 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info