BENNING PV 1-1
19
the BENNING PV 1-1.
- Switch on the BENNING CC 3 current clamp adapter and select the 40 A range.
- Switch on the BENNING PV 1-1 and press the
Auto
-key
3
until the symbol B for current
clamp measurement is displayed.
- For direct current (DC) measurements, press the null balance key (ZERO) of the
BENNING CC 3 until a current value of approx. 0 A is displayed.
- Enclose the single-wire live conductor by means of the current clamp adapter.
- The measured current will be shown on the display
1
.
See gure 8: AC/ DC current measurement by means of optional current clamp adapter
BENNING CC 3
8.6 Measurement on several identical PV modules/ PV strings
For PV systems with several identical PV strings, the open-circuit voltages and short-circuit cur-
rents must be compared with each other at stable radiation conditions. The values should be
identical and should have a maximum deviation of 5 % (at stable radiation conditions). Moreover,
the measured values have to be compared with the values to be expected.
- At the beginning of the series of measurements, press and hold the -key
8
and the
-key
6
in order to delete the measured value memory.
- Connect the first PV string of the PV generator to the BENNING PV 1-1.
- Press the
Auto
-key
3
to start the automatic measurement of the first PV string.
- Press the -key
8
to store the measured values.
- Connect the next PV string to the BENNING PV 1-1 and press the
Auto
-key
3
to start the
automatic measurement.
- Press the -key
8
to store the measured values.
- Repeat this procedure for all identical PV strings at stable radiation conditions.
- After measurement, the stored measured values can be recalled pressing the -key
6
. As
soon as a measured value deviates by more than 5 % from the average value of all stored
measured values regarding the open-circuit voltage or short-circuit current, the ">5%" sym-
bol O starts flashing.
- To call the average values, press the -key
6
repeatedly until the storage location 0 is
displayed. The average values of the open-circuit voltage and of the short-circuit current are
shown on the display
1
.
See gure 6: Automatic solar panel measurement, ISO measurement via earthing conductor
8.7 Measured value memory
8.7.1 Storing measured values
The BENNING PV 1-1 can store up to 200 display screens. Depending on the measure-
ments carried out, the values for protective conductor resistance, open-circuit voltage, short-
circuit current, insulating resistance with testing voltage and even the measured AC/DC current
( BENNING CC 3) are stored with a date/ time stamp.
- Press the -key
8
to store the displayed measured values to the first free storage location.
Successful storage will be confirmed by the "STORE" symbol L on the LC display
1
.
8.7.2 Calling measured values
- Press the -key
6
to recall the stored measured values with the corresponding storage
location number N. The symbol “RECALL” M is displayed
1
. The storage location number
N is shown on the display
1
.
- Press the -key
6
to go to the previous storage location.
8.7.3 Deleting the measured value memory
- To delete the entire measured value memory, press the
-
key
8
and the -key
6
.
8.7.4 Warning in case of the measured value deviating by 5 %
- In case the displayed measured value memory (open-circuit voltage and short-circuit cur-
rent) deviates by 5 % from the average value of all stored measured values, the "5%" symbol
starts flashing.
8.7.5 Average value of the entire measured value memory regarding the open-circuit voltage
and short-circuit current
- Press the -key
6
repeatedly until the storage location 0 N is shown on the display
1
.
The symbol “RECALL” M is displayed
1
.
- The storage location 0 N calculates the average values for open-circuit voltage and short-
circuit current from all stored measured values.
8.7.6 Reading out the measured value memory via the USB interface
To read out the measured values via the USB interface
7
, you have to install the hardware driver
from the "Treiber-driver" directory on the enclosed CD-ROM and then the download program
from the CD-ROM's "Programm-program" directory on your PC once.
In order to download the data, please proceed as follows:
- Disconnect all safety measuring leads from the BENNING PV 1-1.
- Connect the BENNING PV 1-1 to your PC by means of the USB connecting cable.
- The hardware driver is installed automatically on a free COM port and confirms that the new
hardware can be used.
25


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning PV 1.1 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning PV 1.1 in the language / languages: English, German as an attachment in your email.

The manual is 10,6 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Benning PV 1.1

Benning PV 1.1 Quick start guide - Dutch, French, Italian, Spanish - 14 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info