4.3 How to measure average illuminance?
Sometimestheaverageilluminanceinaroommustbefoundout.
A luxmeter indicates the illuminance at the point of measurement only, and not
theaverageilluminanceinthespace.Tofindouttheaverageilluminanceinan
areaatthetime,itisnecessarytodividetheareaintoanumberofequalareas
which should be as nearly square as possible. The illuminance at the centre
ofeachsquareisthenmeasured,andtheresultsareaveraged.Theminimum
number of equal areas required for accuracy can be determined by number n:
L x B
n = -------------------------
(L + B) x H
m
where L is the length, W is the width of the indoor place H
m
and is the height of
theluminariesabovetheplaneofmeasurement.
The height of working plane is usually 0.85 m for work benches or 0.72 m for
desk top height unless the main plane of the work is known to be some other
heightabovefloorlevel.Iftheworkisperformeddowntofloorlevel,thenthe
floor is taken as the working plane of measurement.
Thenumberofmeasurementpointsdependsonvaluen.
n
Minimum number of
measurement points for
uncertainty of ± 5 %
Minimum number of
measurement points for
uncertainty of ± 10 %
n < 1 8 4
1≤n<2 18 9
2≤n<3 32 16
3 < n 50 25
Table 2: Number of measurement points
4.4 Calibration of illumination testers
As ageneral rulesilicon diodetypemeter shouldbe re-calibratedeverytwo
years. Companies conforming to the ISO 9000 standards should have their
meters calibrated every 12 months. It is recommended to recalibrate probe
frequently to maintain specified accuracy.
24


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning LUXMETER Type B at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning LUXMETER Type B in the language / languages: English, German as an attachment in your email.

The manual is 0,83 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info