Fault loop resistance
fault loop resistance both fault loops, R1 (L1-PE) and R2 (L2-PE)
prospective fault current I
SC1
and I
SC2
for both fault loops
Voltage, frequency symbols adapted to a power supply network with lower
voltage
Phase shift automatic recognition of three-phase mains
FI/RCD functions
contact voltage U
c
for both options, U1 (L1-PE) and U2 (L2-PE)
tripping time
maximum nominal differential current is limited to 1 Atripping current
automatic testing
Resistance to earth independent of the selected power supply network
PE conductor test key (TEST key) deactivated
12.3.1 Voltage
Figure 12.2: Example of voltage and frequency measurements
Displayed results for a single-phase system:
L1-L2 voltage between external conductors,
L1-PE voltage between external conductor 1 and protective conductor,
L2-PE voltage between external conductor 2 and protective conductor.
12.3.2 FI/RCD testing
The normal maximum FI/RCD testing current is 1 A RMS (peak current 1.4 A) and can only be
reached,ifthefaultloopresistanceisbelow1Ω.
Tests are carried out automatically for both combinations (L1-PE and L2-PE).
For each individual test result, the corresponding indication is shown.
Figure 12.3: FI/RCD contact voltage test
12.3.3 Line resistance and prospective short-circuit current
The measured resistance represents the phase-to-phase resistance (R
L1-L2
). The nominal voltage
for calculating I
PSC
is set to 110 V.
The nominal voltage range for measuring the line resistance is 90 V to 121 V. If the input voltage
is outside this range, this is displayed on the online voltage/ terminal monitor together with the
symbol.
75


Need help? Post your question in this forum.

Forumrules
1

Forum

benning-it-120b

Reset search

  • when mesuring lightihg earth protection instrument show that is not cinected. why.
    Submitted on 12-8-2023 at 22:36

    Reply Report abuse


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning IT 120B at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning IT 120B in the language / languages: English as an attachment in your email.

The manual is 6,86 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Benning IT 120B

Benning IT 120B User Manual - German - 83 pages

Benning IT 120B User Manual - Dutch - 58 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info