Note:
- If voltage between test terminals is higher than 10 V the continuity measurement cannot be
performed.
- Before performing continuity measurement compensate test lead resistance if necessary. The
compensation is performed in LowΩfunction.
5.3 Testing RCDs (FI/RCD)
When testing RCDs, the following sub-functions can be performed:
- Contact voltage measurement
- Trip-out time measurement
- Trip-out current measurement
- RCD autotest
In general the following parameters and limits can be set when testing RCDs:
- Limit contact voltage
- Nominal differential RCD trip-out current
- Multiplier of nominal differential RCD trip-out current
- RCD type
- Test current starting polarity
5.3.1 Limit contact voltage
Safety contact voltage is limited to 50 V AC for standard domestic area. In special environments
(hospitals, wet places, etc.) contact voltages up to 25 V AC are permitted.
Limit contact voltage can be set in UC Contact voltage function only!
5.3.2 Nominal differential trip-out current
Nominal residual current is the rated trip-out current of the RCD. The following RCD current ratings
can be set: 10 mA, 30 mA, 100 mA, 300 mA, 500 mA and 1000 mA.
5.3.3 Multiplier of nominal residual current
Selected nominal differential current can be multiplied by ½; 1; 2 or 5.
5.3.4 RCD type and test current starting polarity
The BENNING IT 110/ BENNING IT 120 B enables testing of general (non-delayed, G) and selec-
tive (time-delayed, S) RCDs, which are suited for:
- Alternating residual current (AC type, marked with symbol)
- Pulsating DC residual current (A and F type, pulsed current-sensitive, marked with symbol)
- Smooth DC residual current (type B and B+, universal current sensitive, marked with the
symbol)
For RCDs of the types AC, A and F, the fault current can be started with the positive half-wave at
0° or with the negative half-wave at 180°.
29


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning IT 110 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning IT 110 in the language / languages: English as an attachment in your email.

The manual is 6,86 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Benning IT 110

Benning IT 110 User Manual - German - 83 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info