577351
13
Zoom out
Zoom in
Previous page
1/60
Next page
07/ 2014
BENNING CFlex 2
9
Distance from the middle Positioning error
25 mm ± (1.0 % of the final measuring range value)
50 mm ± (2.0 % of the final measuring range value)
75 mm ± (3.0 % of the final measuring range value)
8. Measuring with the BENNING CFlex 2
8.1 Preparing the measurement
Operate and store the BENNING CFlex 2 at the specified storage and operating temperatures
only! Do not permanently expose the device to sunlight.
- Nominal voltage and nominal current of the enclosed safety measuring lead comply with the
respective values of the BENNING CFlex 2. The safety measuring lead is firmly connected
with the BENNING CFlex 2 and is not detachable.
- Check the insulation of the safety measuring lead. If the insulation is damaged, the
BENNING CFlex 2 must be scrapped immediately.
- Strong sources of interference in the vicinity of the BENNING CFlex 2 might involve unsta-
ble readings and measuring errors.
Do not exceed the maximum permitted voltage with respect to earth potential!
Electrical danger!
The highest voltage that may be applied to the BENNING CFlex 2 with respect to earth potential
is 600 V CAT IV / 1000 V CAT III.
8.2 AC current measurement
- Select the measuring range (30 A, 300 A or 3000 A) by means of the rotary switch
.
- Clamp the single-wire live conductor centrally by means of the flexible measuring loop
.
- Read the measured value from the digital display
7
.
See Figure 2 a: Alternating current measurement
See Figure 2 b: Alternating current measurement
9. Maintenance
Before opening the BENNING CFlex 2, make sure that it is free of voltage!
Electrical danger!
Work on the opened BENNING CFlex 2 under voltage may be carried out only by skilled elec-
tricians with special precautions for the prevention of accidents.
Make sure that the BENNING CFlex 2 is free of voltage as described below before opening
the instrument:
- First, remove the BENNING CFlex 2 from the object to be measured.
The BENNING CFlex 2 AC current transformer is not equipped with a fuse.
9.1 Securing the instrument
Under certain circumstances safe operation of the BENNING CFlex 2 is no longer ensured, for
example in the case of:
- Visible damage of the casing.
- Incorrect measurement results.
- Recognisable consequences of prolonged storage under improper conditions.
- Recognisable consequences of extraordinary transportation stress.
In such cases, immediately disconnect the BENNING CFlex 2 from the measuring point and
secure it against further use.
9.2 Cleaning
Clean the exterior of the housing with a clean dry cloth (exception: special cleaning wipers). Avoid
using solvents and/ or scouring agents for cleaning the instrument. It is important to make sure
that the battery compartment and battery contacts are not contaminated by leaking electrolyte.
If electrolyte contamination or white deposits occur in the area of the batteries or battery com-
partment, clean them too with a dry cloth.
9.3 Battery replacement
Before opening the BENNING CFlex 2, make sure that it is free of voltage!
Electrical danger!
13


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning CFlex 2 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning CFlex 2 in the language / languages: English, German, Dutch, French, Italian, Swedish, Turkish, Spanish as an attachment in your email.

The manual is 1,57 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info