6. Environment conditions:
- The BENNING CC 1 is intended for making measurements in dry
environment.
- Maximum barometric elevation for making measurements: 2000 m,
- Overvoltage category IEC 60664/ IEC 61010, 300 V category III,
600 V category II,
- Contamination degree II according to EN 61010-1
- Protection class: IP 30 (EN 60529),
- Operating temperature and relative humidity:
For operating temperatures from 0 °C to 30 °C: relative humidity less than 80
%
For operating temperatures from 31 °C to 40 °C: relative humidity less than 75
%
For operating temperature from 41 °C to 50 °C: relative humidity less than 45 %
- Storage temperature: The BENNING CC 1 can be stored at any temperature
in the range from 20 °C to + 60 °C ).
7. Electrical specications
Note: The measuring precision is specied as the sum of
- a relative fraction of the measured value and
- a current value in A.
The measurement precision applies at a temperature of 23 °C ± 5 °C and a
relative humidity less than 75 %.
7.1 Alternating current measurement
Measurement
range
Measured
value
Output Measurement precision
400 A
3 A 3 mV
± (1,9 % + 0,5 A)
at 50 Hz - 60 Hz
30 A 30 mV
350 A 350 mV
400 A 400 mV
± (3,2 % + 1 A)
at 50 Hz - 60 Hz
Max. equivalent source resistance: 75 Ω
8. Making measurements with the BENNING CC 1
8.1 Preparations for making measurements
Operate and store the BENNING CC 1 only at the specied storage and
operating temperatures. Avoid continuous insulation.
- The nominal voltage and current ratings of the safety measuring cables
included in the scope of delivery correspond to the ratings of the
BENNING CC 1.
- The safety measuring cable is securely connected to the BENNING CC 1
and cannot be removed.
- Check the insulation of the safety measuring cables. Discard the
BENNING CC 1 immediately if the insulation is damaged.1
- Do not clamp the measuring pliers on a live conductor before the
BENNING CC 1 has not been connected to a multimeter.
Strong sources of interference in the vicinity of the BENNING CC 1 can lead
to unstable readings and measuring errors.
- Do not apply any voltage to the output contacts of the BENNING CC 1.
Do not exceed the maximum permitted voltage with respect to
earth potential! Electrical danger!
The highest voltage, that may be applied to the BENNING CC 1 against ground
potential, is 600 V.
8.2 Alternating current measurement
- Set the multimeter to the function “alternating voltage measurement” (V AC)
and select a measurement range that can display voltages from 1 mV to
400 mV.
- Contact the 4 mm safety connector of the spiral type safety measuring cable
with the COM jack of the multimeter.
- Contact the red 4 mm safety connector of the spiral type safety measuring
cable with the voltage input jack (V) of the multimeter.
- Opening lever
Operate and use meas. pliers to
clamp the single wire
live conductor.
- Read off the voltage value on the multimeter and convert to the current
value taking into account the conversion factor.
cc_1_in.indd 24.09.2003, 10:137
9


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning CC1 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning CC1 in the language / languages: English, German, Dutch, Swedish, Spanish as an attachment in your email.

The manual is 0,42 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info