Electromagnetic Compatibility (EMC, IEC 60601-1-2, Table 2)
Electromagnetic Immunity
Calypso is intended for use in the electromagnetic environment specified below.
The customer or the user of Calypso should assure that it is used in such environment.
Immunity Tests IEC 60601 test level Compliance level Electromagnetic environment
- guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV
for power supply lines
± 1kV
for input / output lines
± 2kV
for power supply lines
n/a
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1kV
line(s) to line(s)
± 2kV
line(s) to earth
± 1kV
line(s) to line(s)
n/a
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
< 5% UT
(> 95% dip in UT)
for 0.5 cycle
< 5% UT
(> 95% dip in UT)
for 0.5 cycle Mains power quality should be
that of a typical commercial or
hospital environment.
If the user of Calypso requires
continued operation during
power mains interruptions, it is
recommended that Calypso be
powered from an uninterruptible
power supply or a battery.
40% UT
(60% dip in UT)
for 5 cycles
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
70% UT
(30% dip in UT)
for 25 cycles
< 5% UT
(> 95% dip in UT)
for 5 sec
< 5% UT
(> 95% dip in UT)
for 5 sec
Power frequency
magnetic field
IEC 61000-4-8
3A/m >30A/m
Power frequency magnetic
fields should be at levels of a
typical commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
68
68


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Ardo Calypso Double Plus at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Ardo Calypso Double Plus in the language / languages: English, German, Dutch, French, Italian, Portuguese, Turkish, Spanish as an attachment in your email.

The manual is 8.39 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Ardo Calypso Double Plus

Ardo Calypso Double Plus User Manual - English, Danish, Polish, Swedish, Norwegian, Finnish - 84 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info