Technical Data
Type UM-201
Measurement method Oscillometric measurement
Measurement range Pressure: 0 - 299 mmHg
Pulse: 40 - 180 beats / minute
Measurement accuracy Pressure: ±3 mmHg
Pulse: ±5 %
Power supply 4 x 1.5V batteries (R6P, LR6 or AA) or
AC adapter (TB-233) (Not included)
Number of measurements
Approx. 450 measurements, when AA Alkaline
batteries are used, with pressure value of 180
mmHg at room temperature of 23°C
Classification Internally powered ME equipment (Supplied by
batteries)
Class II (Supplied by adapter)
Continuous operation mode
Clinical test According to ANSI / AAMI SP-10 1992
EMC IEC 60601-1-2: 2007
Operating conditions +10°C to +40°C / 15 %RH to 85 %RH
800 hPa to 1060 hPa
Transport / Storage conditions
-10°C to +60°C / 15 %RH to 85 %RH
Dimensions Approx. 150 [W] x 156 [H] x 120 [D] mm
Weight Approx. 480 g, excluding the batteries
Applied part Cuff Type BF
Useful life Device: 5 years (when used six times a day)
Cuff: 2 years (when used six times a day)
Accessory AC adapter The adapter is to connect the blood pressure
monitor to a power source at home.
TB-233 Please contact your local A&D dealer for
purchasing.
The AC adapter is required to be inspected or
replaced periodically.
TB-233C Input: 100-240V
Output: 6V
500mA
TB-233BF Input: 100-240V
Output: 6V
500mA
139°C 2A
139°C
2A
16


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for AnD UM-201 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of AnD UM-201 in the language / languages: English as an attachment in your email.

The manual is 0,31 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info