Model UA-1020 is designed to detect the pulse and to inflate the cuff to a
systolic pressure level automatically.
If re-inflation occurs repeatedly, use the following methods.
Measurement with the SET Pressure
During the blood pressure measurement, re-inflation may occur.
A fixed pressure value can be set to avoid re-inflation.
1. At step 6 of the clock adjustment procedure
(see page 9), press the SET button to go to
the pressure setting mode. The current
setting blinks.
2. Press the or button to select a pressure
value about 30 mmHg or more above your
expected systolic pressure from the following.
AUTO : Automatic pressurization (default value)
180 : Pressure value of 180 mmHg (fixed)
210 : Pressure value of 210 mmHg (fixed)
240 : Pressure value of 240 mmHg (fixed)
3. Press the SET button to go to the TriCheck setting mode.
Press the START button to turn the device off. After three minutes of
non-operation, the device will turn off automatically.
The next measurement will be performed with the new pressure value.
Measurement with the Desired Systolic Pressure
Use this method when re-inflation occurs repeatedly even if the pressure value
is set to 240 in the procedure above or when the results are not displayed
even if the pressure decreases to 20 mmHg or less.
1. Place the cuff on the arm (preferably the left arm).
2. Press and hold the START button until a
number about 30 to 40 mmHg higher
than your expected systolic pressure
appears.
3. When the desired number is reached,
release the
START button to start
measurement. Continue to measure
your blood pressure as described on
the previous page.
START
Refer to the previous page
for measurement
Press and hold the button
At heart level
Release the button
at the desired
systolic pressure
or
13


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for AnD UA-1020 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of AnD UA-1020 in the language / languages: English as an attachment in your email.

The manual is 0,6 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info