Table 10. Guidance and manufacturer’s declaration – electromagnetic immunity for
the Aerogen Solo nebuliser system that is not life supporting
This Aerogen Solo nebuliser system is intended for use in the electromagnetic
environment specified below. The customer or the user of the Aerogen Solo nebuliser
system should assure that it is used in such an environment.
Immunity Test IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment – Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
EN 61000-4-2:
2009
±8 kV contact
±15 kV air
±2, 4, 6 & 8 kV
contact
±2, 4, 6, 8 &
15 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast
Transient/burst
IEC 61000-4-4
EN 61000-4-4:
2012
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
EN 61000-4-5:
2006
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
EN 61000-4-11:
2004
<5 % Ut
(>95 % dip in Ut)
for 0.5 cycle @
0°, 45°, 90°, 135°,
180°, 225°, 270°,
315°
70 % Ut
(30 % dip in Ut)
for 25 cycles
<5 % Ut
(>95 % dip in Ut)
for 5 sec
<5 % Ut
(>95 % dip in Ut)
for 0.5 cycle @
0°, 45°, 90°, 135°,
180°, 225°, 270°,
315°
70 % Ut
(30 % dip in Ut)
for 25 cycles
<5 % Ut
(>95 % dip in Ut)
for 5 sec
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the Aerogen Solo nebuliser
system requires continued
operation during power mains
operation, it is recommended
that the Aerogen Solo
nebuliser system be powered
from an uninterruptible power
supply or battery.
40


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Aerogen Solo at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Aerogen Solo in the language / languages: English as an attachment in your email.

The manual is 5.4 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Aerogen Solo

Aerogen Solo User Manual - German - 39 pages

Aerogen Solo User Manual - Dutch - 42 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info