Aerogen® Solo System Instruction Manual 31
Performance
Table 6. Performance Specifications of the Aerogen Solo
Flow Rate > 0.2 mL/min (Average ~ 0.38 mL/min)
Particle Size
As measured with the Andersen Cascade Impactor:
Specification Range: 1-5 μm
Average Tested: 3.1 μm
As measured with the Marple 298 Cascade Impactor:
Specification Range: 1.5-6.2 μm
Average Tested: 3.9 μm
As per EN 13544‑1:
Aerosol Output rate: 0.30 mL/min
Aerosol Output: 1.02 mL emitted of 2.0 mL dose
Residual Volume: <0.1 mL for 3 mL dose
Performance may vary depending upon the type of drug and nebuliser used. For
additional information contact Aerogen or drug supplier.
The temperature of the medication will not rise more than 10°C (18°F) above ambient
during normal use.
Representative particle size distribution for Albuterol as per EN 13544-1 is
shown below.
0%
100%
0.1 1 10 100
50%
Particle Size (μm)
Cumulative Undersize %
33


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Aerogen Solo at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Aerogen Solo in the language / languages: English as an attachment in your email.

The manual is 5.4 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Aerogen Solo

Aerogen Solo User Manual - German - 39 pages

Aerogen Solo User Manual - Dutch - 42 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info