3. Repress the prism cover [1]. The liquid quantity
should be measured in such a way that most of
the prism surface is moistened.
4. Look through the ocular [4] and point the refrac-
tometer including its prism surface [2] towards a
bright light source if possible.
5. Looking through the ocular [4], you will see a
light and dark field. Its boundary lines indicate
the refractive index on the scale also visible in
the ocular [4].
6. By using the calibration tools [7], the scale can
be adjusted and the measuring device calibrated
by turning the calibration screw [3] behind the
prism surface [2] (underneath the rubber cap).
The light-dark-border line must coincide with the
waterline. The room temperature should be at
+20 °C.
Caution!
The scales are set at an ambient air tem-
perature of +20 °C!
Measuring Process for Antifreeze
Protection-Check
1. Lift the prism cover [1] and apply a few drops of
cooling solution to be measured to the prism sur-
face [2] using the included pipette [10]. Distribute
the liquid evenly by pressing the prism cover [1].
2. Close the prism cover [1] and look at the measu-
ring scale through the ocular [4]. Point the refrac-
tometer including its prism surface [2] towards a
bright light source if possible.
3. Turn the adjustment ring [5] on the ocular [4] to
set the focus of the vision.
4. The border line on the measuring scale moves
depending on the anti-freeze agent concentra-
tion. This light-dark-border directly shows the
freezing point of the anti-freeze mix in °C on
the anti-freeze-scale [B]. The values show the
measured value for Propylene* (G13) and the
measured value for Ethylene** (G11 / G12). The
Propylene scale must also be used if you are not
familiar with the poured-in anti-freeze protection
or if it is a mixture of both. This is the only way
sufficient anti-freeze protection of the cooling
17


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for AEG RX 3 - 97200 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of AEG RX 3 - 97200 in the language / languages: English, German, French, Italian as an attachment in your email.

The manual is 3,62 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info