Basic use and operation
Setup
The AV500 needs to be placed in the charging cradle before
rst use. AccuVein recommends fully charging the AV500 at
that time. Connect the charging cradle to the power supply.
The AV500 comes with power supply adaptors to ensure outlet
compatibility. If necessary, secure a dierent power adaptor
to the power supply before plugging it into an outlet. The
AV500 is compatible with all voltages from 100
240VAC.
Remove the plastic protective cover from the AV500 display
screen before use.
Intended Use
The AccuVein AV500 is a portable, hand-held instrument
that helps medical professionals to locate certain supercial
veins. The AV500 is intended to be used as a supplement
to appropriate medical training and experience. The AV500
should not be used as the sole method for locating veins,
and should be used only by a qualied medical professional,
who should do so either prior to palpation to help identify
the location of a vein, or afterwards to conrm or refute the
perceived location of a vein. When using the AV500, medical
practitioners should always follow the appropriate medical
protocols and practices as required by their medical facility,
as well as exercise sound medical judgment.
When used properly, the AV500 enables users to locate certain
supercial veins in connection with medical procedures, such
as venipuncture.
The AV500 can be used whenever the determination of vein
location is appropriate such as hospitals and clinics.
Product Description
The AV500 operates by using infrared light to detect veins
beneath the skin, then projecting the position of the veins
on the skin surface directly above the veins. Qualied medi-
cal personnel can observe the vasculature as displayed to
assist them in nding a vein of the right size and position
for venipuncture and other medical procedures requiring
the location of supercial veins. No training is required to
operate the AV500.
The AV500 only shows supercial vasculature. The maximum
depth that veins are displayed varies by patient. In addition,
some patients veins or a portion of their veins might not be
displayed well or at all. Causes for less than optimal or lack
of vein display include, but are not limited to, vein depth,
skin conditions (e.g., eczema, tattoos), hair, scarring or other
highly contoured skin surface, and adipose (fatty) tissue.
When held directly overhead, the AV500 accurately locates
the center of a vein. Increasing the displacement from directly
overhead results in an oset in the displayed vein position.
Width of displayed vein may dier from the actual width
depending on patient to patient dierences and vein width.
The center line of the vein is accurate when the device is
being used correctly and should always be used as the target
when performing venipuncture or other medical procedures.
The AV500 requires no routine or preventative maintenance.
The AV500 is portable, internally powered by battery and
approved for continuous operation. The AV500 is considered
a Type B applied part.
The technology in the AV500 is patented. Visit www.accuvein.com
for the most current information.
5
5


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for AccuVein AV500 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of AccuVein AV500 in the language / languages: English as an attachment in your email.

The manual is 1.23 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of AccuVein AV500

AccuVein AV500 User Manual - German - 16 pages

AccuVein AV500 User Manual - Dutch - 16 pages

AccuVein AV500 User Manual - French - 16 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info