The standard power-on display test checks the most important
display elements. To verify that all of the elements are functioning
correctly, you can carry out a full test.
씰 Press and hold down the M and S buttons together for longer
than 3 seconds.
씰 All the elements of the liquid crystal display (LCD) are shown
at once. Either “mmol/L” or “mg/dL” will be visible depending
on the country-specific setting.
씰 Press any key to terminate the display test and turn off your
Accu-Chek Active meter. If you do not press a key, the meter
will shut off automatically after about 2 minutes.
97
man_04493168001_02_EnUK.indd 97 1/12/2006, 11:29:16 ðì