12. SW1 press switch (picture 21)
ON/OFF switch
The SW1 switch acts as the on/off switch if this function is activated (DIP switch 6 is
down).
Press the SW1 switch for three (3) seconds in order to activate or deactivate the control
unit. When the control unit is switched on, you will hear a beep and the blue LED light will
start to flash. When the device is switched off, you will hear a "beep-beep-boop" (high-
highlow) sound. The blue LED light will go off.
Acknowledgement switch
The notification sound (if activated) can be muted by pressing
the SW1 switch shortly. The device will make a "beep-beep"
sound.
NOTE! If you press the switch for too long, you might
accidentally switch off the device. The notification sound will
also stop when the fast-paced movements stop.
Bed exit alarm “presence / absence” sensitivity switch
SW1 switch can be used to launch automatic calibration for presence / absence sensitivity.
See chapter 13. Calibrating the “presence / absence” sensitivity.
13. Calibrating the “presence / absence” sensitivity
Doing calibration
The sensitivity to notice correctly that person is in or out off bed, is adjusted in calibration
mode. First the person to be monitored should be in bed and resting (not moving, talking
or else) at least for one (1) minute until green light starts to blink with same speed as blue
light. Control unit should lay on a table or be fixed on a wall. Sensor must be in its correct
place under mattress.
Start calibration by pressing three (3) times the SW1 switch (picture 21). You will hear 3
times low and high tone (kind of bu-beep, bu-beep, bu-beep). The calibration takes 18
seconds. If calibration was successful, the confirmation sound is the same 3 times bu-
beep again.
Error messages and how to handle them
One (1) long beep - Device has not detected strong enough
micro movement signal and has reached an adjustment limit.
Device might not work optimally. Is the sensor positioned
correctly in the bed and device was not in error -state (red led
was not blinking)? If not: Correct the problem and re-try
calibration. If yes: Adjust the rotary switch (picture 5) up 2
notches and re-try auto adjustment.
Two (2) long beeps - Device has detected too strong signal and
has reached an adjustment limit. Device might not work optimally. Was person laying still
during the calibration period? Re-try calibration again and see if second time would be
succesful. If second time gives same error sound, adjust the rotary switch (picture 5) down
2 notches and do the calibration again.
17


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Abilia 464032 Emfit SafeBed at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Abilia 464032 Emfit SafeBed in the language / languages: English as an attachment in your email.

The manual is 1,11 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info