677197
19
Zoom out
Zoom in
Previous page
1/36
Next page
19
14. Signal sounds
There are two types of signal sounds.
Alarm signals, that are long and loud, are used to notify about fast movements and bed
exit. Alarm signals can be muted using the DIP switches (see chapter 6.4 Adjusting the
notification sound volume).
Information signals, that are short and quieter than alarm signals, are used as feedback to
the user. Information signals tell the user when the device is turned on or off or when
alarm signal is acknowledged. Possible malfunctions of the equipment (disconnected bed
sensor, empty batteries) and feedback about calibration of the sensor are also notified with
information signals. Information signals can’t be muted.
Alarm signals are high priority signals that will overrule information signals if both signals
activate simultaneously.
14.1 Alarm signals
Two high and low “BEEPS” in a loop.
Fast movements alarm goes off.
Three high and low “BEEPS” in a loop.
Bed exit alarm goes off.
14.2 Information signals
Short, high “BEEP”.
SW1 is pressed and device turns on.
Short, low “BEEP”.
SW1 is pressed and device is turned off.
Short, low “BEEP” x 2.
SW1 has been pressed once and alarm signal
is acknowledged.
Short, high “BEEP” x 3 and short low
“BEEP” x 3.
Sensor is disconnected or broken. Information
signal sounds after 10 seconds of
disconnection and then every 45 seconds.
Short, high “BEEP”, short “LOWbeep.
Batteries are empty. Replace batteries.
Short, high “BU-BEEP” x 3.
Sensitivity calibration has started / successful
calibration.
Short, low “BEEP” (after calibration).
Sensitivity too low.
Short, low “BEEP” (after calibration) x 2.
Sensitivity too high.
19


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Abilia 464031 Emfit Monitor at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Abilia 464031 Emfit Monitor in the language / languages: English as an attachment in your email.

The manual is 1,09 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Abilia 464031 Emfit Monitor

Abilia 464031 Emfit Monitor User Manual - German - 36 pages

Abilia 464031 Emfit Monitor User Manual - Dutch - 36 pages

Abilia 464031 Emfit Monitor User Manual - French - 32 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info