8. Connectors and wiring
Connect the sensor, power supply (sold separately) and any additional cables in the
following way:
Connect the sensor to
input X3
It’s possible to attach another
system to the control unit via
input X2 (i.e. Nurse call system)
The power supply (sold
separately) fits into input
X1.
Connectors are marked with X1, X2 and X3. These are:
X1 Connector for an external power supply. Only
use Globtek inc. power supply (see chapter Technical
Specifications for details) which can be obtained as an
original accessory from Abilia.
X2 AUX Connector to connect the device to a nurse call system, a personal
emergency phone or an external wireless transmitter. The connector may only be
connected to a system safety voltage input with max. voltage below 25 V(AC) / 60 V (DC),
where both poles have been separated from the electrical network. Max. load current
100 mA.
X3 - Sensor connector. (use only Emfit floor sensor model L-6090SL or L-60150)
8.1 X2 (AUX) Connector Pin Order
From left to right:
Pin 1
Common return
Pin 2
Normally Open (NO) send
Pin 3
Normally Closed (NC) send
Pin 4
Low Battery warning send (NO)
Pin 5
Not in use, do not connect
Pin 6
Not in use, do not connect
Pin 7
Not in use, do not connect
Pin 8
Not in use, do not connect
Dry-contact output of the X2 (AUX) connector can only be connect to a system
safety voltage input with max voltage below 25V (AC) / 60V (DC), where both
poles have been separated from the electrical network (so called floating applied
part)! Max load current is 100mA.
11


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Abilia 464020 - Emfit SafeFloor at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Abilia 464020 - Emfit SafeFloor in the language / languages: English as an attachment in your email.

The manual is 1,03 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info