10
6. Setting up the dip switches
Open the lid of the control unit by lifting it from the side (picture 9). Inside, you will find
eight DIP switches that are used to select the desired functions (picture 4). Remember to
select the desired settings using the DIP switches before using the device.
The control unit has the following factory settings. DIP 1, 2, 3, 4, 5, 6, 7 and 8 OFF (down).
With these settings:
• Bed exit notification delay is 3 seconds.
• Output is in pulse mode
• The SW1 switch acts as the On/Off switch which means that the device can be
switched on or off by pressing the switch for three (3) seconds.
• The notification sound is “very loud”.
Disconnect the device from the power supply and remove nurse call cable (if
used) from connector X2 before opening the lid.
Remove the batteries and disconnect the power supply before setting up the DIP
switches in order to implement the new settings.