24. Electromagnetic conditions
System specification:
- D-1090-2G monitor
- L-4060SL or L-4060SLC bed sensor
- GlobTek external power supply (see chapter Technical Specifications for details)
Cable specification:
- Power cable (non-shielded) max. Length 2 m
- Sensor cable (shielded) max. length 3 m
Note! RF communications equipment can effect medical electrical equipment!
Guidance and manufacturer’s declaration – electromagnetic emissions
The Emfit Tonic-Clonic Seizure Monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of the equipment should
assure that it is used in such an environment.
RF emissions CISPR 11
Group 1
The Emfit Tonic-Clonic Seizure Monitor uses
RF energy only for its internal function.
Therefore, its RF emissions are very low and
are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11
Class B
The Emfit Tonic-Clonic Seizure Monitor is
suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions IEC
61000-3-2
Class A
Voltage fluctuations/
flicker emissions IEC
61000-3-3
Complies
28


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Abilia 464000 Emfit Monitor at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Abilia 464000 Emfit Monitor in the language / languages: English as an attachment in your email.

The manual is 1,09 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Abilia 464000 Emfit Monitor

Abilia 464000 Emfit Monitor User Manual - German - 36 pages

Abilia 464000 Emfit Monitor User Manual - Dutch - 36 pages

Abilia 464000 Emfit Monitor User Manual - French - 32 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info